Dew point set DS 400 for monitoring of medical compressed air

The monograph "Aer medicalis" in the supplement to the European Pharmacopoeia contains - in comparison to DIN EN 737-3 - significantly changed purity characteristics, whereby in particular the drastic lowering of the dew point temperature to -46°C atmospheric (this corresponds to 67 ppm V/V) leads to considerable difficulties. The value 67 ppm V/V for water corresponds to the condensation limit at a Pressure of 200 bar and a Temperature of 11 ºC. This limitation of the water content to the condensation limit is necessary because the condensation of water can cause corrosion damage and also favour the proliferation of germs.

to the inquiry

The product has been added to inquiry.
to the inquiry

Details

  • Chart recorder DS 400
  • Dew point sensor FA 510
  • Standard measuring chamber
  • Pressure sensors CS 10

Related products

DS 400 Set - Stationary dew point measurement in compressed air systems

Ideal for Dew point measurement of Refrigeration dryers and Desiccant dryers.

More

FA 510/515 - Dew point sensor for adsorption dryers

For residual moisture measurement from -80 to 20°Ctd. New: With Modbus RTU interface.

More

Standard measuring chamber

For compressed air up to 16 bar

More

Standard pressure sensor CS 10 and CS 16

for Measuring 0...10 and 0...16 bar with an Accuracy of max. 1% or max. 0.5% deviation.